Fluorescence
Fluorescent probes enable researchers to detect particular components
of complex biomolecular assemblies, including live cells, with exquisite
sensitivity and selectivity. Fluorescence is the result of a three-stage
process, see below, that occurs in certain molecules (generally polyaromatic
hydrocarbons or heterocycles) called fluorophores or fluorescent dyes.
A fluorescent probe is a fluorophore designed to localize within a specific
region of a biological specimen or to respond to a specific stimulus.
Stage 1 : Excitation
A photon of energy hvEX is supplied
by an external source such as an incandescent lamp or a laser and absorbed
by the fluorophore, creating an excited electronic singlet state
(S1').
Stage 2 : Excited-State Lifetime
The excited state exists for a finite time (typically 1–10 nanoseconds).
During this time, the fluorophore undergoes conformational changes and
is also subject to a multitude of possible interactions with its molecular
environment. These processes have two important consequences. First, the
energy of S1' is partially dissipated,
yielding a relaxed singlet excited state (S1)
from which fluorescence emission originates. Second, not all the molecules
initially excited by absorption (Stage 1) return to the ground state
(S0)
by fluorescence emission. Other processes such as collisional quenching,
fluorescence energy transfer and intersystem crossing may also depopulate
S1. The fluorescence quantum yield, which
is the ratio of the number of fluorescence photons emitted (Stage 3) to
the number of photons absorbed (Stage 1), is a measure of the relative
extent to which these processes occur.
Stage 3 : Fluorescence Emission
A photon of energy hvEM is emitted,
returning the fluorophore to its ground state
S0. Due to energy dissipation during the
excited-state lifetime, the energy of this photon is lower, and therefore of
longer wavelength, than the excitation photon
hvEX. The difference in energy
or wavelength represented by (hvEX
–hvEM) is called the Stokes shift.
The Stokes shift is fundamental to the sensitivity of fluorescence techniques
because it allows emission photons to be detected against a low background,
isolated from excitation photons. In contrast, absorption spectrophotometry
requires measurement of transmitted light relative to high incident light
levels at the same wavelength.
Fluorescence Spectra
The entire fluorescence process is cyclical. Unless the fluorophore is
irreversibly destroyed in the excited state (an important phenomenon known
as photobleaching), the same fluorophore can be repeatedly excited and
detected. For polyatomic molecules in solution, the discrete electronic
transitions represented by hvEX
and hvEM are replaced by rather
broad energy spectra called the fluorescence excitation spectrum and
fluorescence emission spectrum, respectively. With few exceptions, the
fluorescence excitation spectrum of a single fluorophore species in dilute
solution is identical to its absorption spectrum. Under the same conditions,
the fluorescence emission spectrum is independent of the excitation wavelength,
due to the partial dissipation of excitation energy during the excited-state
lifetime. The emission intensity is proportional to the amplitude of the
fluorescence excitation spectrum at the excitation wavelength.
Fluorescence Signals
Fluorescence intensity is quantitatively dependent on the same parameters
as absorbance — defined by the Beer–Lambert law as the product of the
molar extinction coefficient, optical pathlength and solute concentration
— as well as on the fluorescence quantum yield of the dye and the excitation
source intensity and fluorescence collection efficiency of the instrument.
In dilute solutions or suspensions, fluorescence intensity is linearly
proportional to these parameters. When sample absorbance exceeds about
0.05 in a 1 cm pathlength, the relationship becomes nonlinear and measurements
may be distorted by artifacts such as self-absorption and the inner-filter
effect. Because fluorescence quantitation is dependent on the instrument,
fluorescent reference standards are essential for calibrating measurements
made at different times or using different instrument configurations.
Background Fluorescence
Fluorescence detection sensitivity is severely compromised by background
signals, which may originate from endogenous sample constituents (referred
to as autofluorescence) or from unbound or nonspecifically bound probes
(referred to as reagent background). Detection of autofluorescence can
be minimized either by selecting filters that reduce the transmission of
E2 relative to E1 or by selecting probes that absorb and emit at longer
wavelengths. Although narrowing the fluorescence detection bandwidth increases
the resolution of E1 and E2, it also compromises the overall fluorescence
intensity detected. Signal distortion caused by autofluorescence of cells,
tissues and biological fluids is most readily minimized by using probes
that can be excited at >500 nm. Furthermore, at longer wavelengths, light
scattering by dense media such as tissues is much reduced, resulting in
greater penetration of the excitation light.
Ratiometric Measurements
In some cases the free and ion-bound forms of fluorescent ion indicators
have different emission or excitation spectra. With this type of indicator,
the ratio of the optical signals (S1 and S2) can be used to monitor the
association equilibrium and to calculate ion concentrations. Ratiometric
measurements eliminate distortions of data caused by photobleaching and
variations in probe loading and retention, as well as by instrumental factors
such as illumination stability.